超低頻開路互補式環形共振器應用於全血凝血時間檢測之設計與實現 Design and Implementation of Ultra-Low Frequency Open Complementary Split-Ring Resonator for Whole Blood Coagulation Time Detection |
| 指導教授: |
楊慶隆
Yang, Chin-Lung |
| 學位類別: |
碩士 Master |
| 系所名稱: |
電機資訊學院 - 電機工程學系 Department of Electrical Engineering |
| 論文出版年: | 2016 |
| 畢業學年度: | 104 |
| 語文別: | 中文 |
| 論文頁數: | 66 |
| 中文關鍵詞: | 開路互補式開口環形共振腔 、凝血酶原時間 、共振器法 |
| 外文關鍵詞: | Open Complementary Split Ring Resonators, Prothrombin Time, Resonator Methods |
| 相關次數: | 點閱:120 下載:16 |
| 分享至: |
| 查詢本校圖書館目錄 查詢臺灣博碩士論文知識加值系統 勘誤回報 |
本文提出一個開路互補式開口環形共振腔(Open Complementary Split Ring Resonator, OCSRR)微波感測器應用於全血凝血時間檢測,為至目前為止第一個首創以開路互補式開口環形共振腔量測凝血變化。藉由擷取網路分析儀S參數的訊號,並經由演算法萃取共振頻率的變化量,而得其介電常數之變化,可以偵測血液凝血時微弱的變化量並準確的量測到血液的凝固時間。不同於目前探測血液凝固的現有技術,吾人希望利用微波共振腔的優點:高靈敏度、少量待測物、容易製造、非接觸式檢測等等優點進行檢測凝血凝固。
OCSRR微波感測器應用於凝血檢測需克服三個主要問題:頻帶選擇、OCSRR微型化、和高損耗待測物。吾人利用阻抗分析儀針對血液進行實驗,藉由柯爾柯爾圖分析血液的特徵頻率,可得知全血凝血最大的變化量約在1 MHz。接著我們以電磁場分布決定最佳之外掛電容與電感被動元件位置,將諧振腔頻率降至血液的特徵頻率以獲得較大的凝血訊號變化。為了解決血液高損耗與凝血變化量不明顯的情形,需先將儀器設置在低雜訊取樣,並以Matlab演算法處理共振腔訊號中的雜訊,將擷取諧振頻以時域的方式呈現,最後藉由一階微分最大值來判斷凝血訊號變化最大的瞬間。本研究以血漿、凝血、不凝血對照實驗進行驗証,分析凝血曲線的變化與趨勢。因紅血球的聚合、紅血球形變、與沉降效應導致總體之等效介電常數改變,進而影響OCSRR諧振腔的共振頻率。最後我們還與光學法比較凝血時間,兩者之間有良好相依性且都有高度的準確性。吾人所使用的平面共振腔法量測的20管血中,手撈法PT與平面共振器量測得PT相關係數為0.862,平均誤差秒數為1.56秒。本研究成功驗証所提出之開路互補式開口環形共振腔,可以簡單使用、成本低廉、直接偵測全血之凝血時間。
This thesis presents applying microwave sensors based on open complementary split ring resonators (OCSRRs) design to measure the blood coagulation time of the whole blood, which is the first time proof that a microwave wave plane resonance method can be effectively applied to detect the blood coagulation time. The S-parameters were recorded and the frequency deviation was computed using interference-suppression algorithm to detect the slight variation of the permittivity during clotting process. Different from the current well-known methods of blood coagulation time detection, the microwave resonator method has several advantages, including high sensitivity, small amount of samples, easy processing, and non-contact measurement for the whole blood coagulation time detection.
There are three major challenges to apply OCSRR microwave sensors for blood coagulation time detection, including the frequency selection, the miniaturization of OCSRR, and highly lossy test sample. We used an impedance analyzer for blood test and analyzed the characteristic frequency of the whole blood. Based on analysis and comparison of the Cole-Cole diagram, the most significant change of the permittivity can be observed near 1 MHz during the coagulation. After analyzing the electric and magnetic field distributions of OCSRR, semi lumped components of capacitors and inductors are used to miniaturize the size of OCSRR to resonate at 1 MHz for the detection of the whole blood property. To solve slight variation of the dielectric constant of the highly lossy material blood, Matlab interference suppression algorithms was applied after a low-noise sampling by the Network Analyzer to filter out the high frequency noise. The captured resonant frequencies were traced in the time domain to observe the change time. Then the prothrombin time (PT) can be measured clearly from the frequency deviation. Finally, the experimental verification including plasma, blood clotting, non-coagulated blood samples was perform, and we can conclude that the aggregation, the shape transformation and the sedimentation of erythrocytes during the coagulation process are the three major factors to result in the permittivity change. Moreover, the resonator methods are compared with the well-established optical methods, both of methods had good accuracy. We had measured data of 20 tubes of whole blood, and the correlation coefficient between PT of resonator methods and PT of manual methods is 0.862. The resonator measuring has the average error of 1.56 sec from the standard manual method.
[1] http://www.hpa.gov.tw/Bhpnet/Web/Index/Index.aspx
[2] 何敏夫, 血液學. 合記出版社, 1993.
[3] S. Niessen, H. Hoover, and A. J. Gale, "Proteomic analysis of the coagulation reaction in plasma and whole blood using PROTOMAP," Proteomics, vol. 11, pp. 2377-88, Jun 2011.
[4] F. Hobbs, B. Delaney, D. Fitzmaurice, S. Wilson, C. Hyde, G. Thorpe, et al., “A review of near patient testing in primary care,” Health Technology Assessment (Winchester, England), vol. 1, pp. i-iv, 1-229, 1996.
[5] M. Kaibara, “Rheology of blood coagulation,” Biorheology, vol. 33, pp. 101-117, 1996.
[6] C. Solano, P. Zerafa, and R. Bird, “A study of atypical APTT derivative curves on the ACL TOP coagulation analyser,” International journal of laboratory hematology, vol. 33, pp. 67-78, 2011.
[7] M. M. Tripathi, Z. Hajjarian, E. M. Van Cott, and S. K. Nadkarni, "Assessing blood coagulation status with laser speckle rheology," Biomedical optics express, vol. 5, pp. 817-831, 2014.
[8] T. P. Vikinge, K. M. Hansson, J. Benesch, K. Johansen, M. Ra, T. L. Lindahl, et al., “Blood plasma coagulation studied by surface plasmon resonance,” Journal of biomedical optics, vol. 5, pp. 51-55, 2000.
[9] J. Machado, A. Lenzi, W. Silva, and R. Sigelmann, “An ultrasonic method to measure human plasma coagulation time,” The Journal of the Acoustical Society of America, vol. 90, pp. 1749-1753, 1991.
[10] T.J. Cheng, H.C. Chang, and T.M. Lin, “A piezoelectric quartz crystal sensor for the determination of coagulation time in plasma and whole blood,” Biosensors and Bioelectronics, vol. 13, pp. 147-156, 1998.
[11] H.C. Chang, T.J. Cheng, T.H. Wu, and T.M. Lin, “Determination of coagulation time in whole blood containing anticoagulant by piezoelectric quartz crystal sensor,” Sensors and Actuators B: Chemical, vol. 66, pp. 296-298, 2000.
[12] H. Lim, J. Nam, S. Xue, and S. Shin, “Measurement of blood coagulation with considering RBC aggregation through a microchip-based light transmission aggregometer,” Clinical hemorheology and microcirculation, vol. 47, pp. 211-218, 2011.
[13] A. Ur, "Determination of blood coagulation using impedance measurements," Biomedical engineering, vol. 5, p. 342, 1970.
[14] 陳家成, 以電阻抗分析法測量人類全血凝固時間, 國立成功大學, 台南, 台灣., 2005.
[15] R. L. Rosenthal and C. W. Tobias, “Measurement of the electric resistance of human blood; use in coagulation studies and cell volume determinations,” J Lab Clin Med, vol. 33, pp. 1110-22, Sep 1948.
[16] H. H. Henstell, I. S. Henstell, and E. M. Ornitz, Jr., “The electrolytic resistance of the blood clot in polycythemia vera before and after radiation and its relationship to clot retraction,” Am J Clin Pathol, vol. 21, pp. 820-7, Sep 1951.
[17] H. Berney and J. O’Riordan, “Impedance measurement monitors blood coagulation,” Analog Dialogue, vol. 42, pp. 42-08, 2008.
[18] J. Connelly and M. Buckler, “The continuous measurement of resistivity and permittivity of human blood plasma during coagulation,” Medical and biological engineering, vol. 13, pp. 523-530, 1975.
[19] Y. Hayashi, M.-A. Brun, K. Machida, and M. Nagasawa, “Principles of Dielectric Blood Coagulometry as a Comprehensive Coagulation Test,” Anal. Chem., vol. 87, pp. 10072-10079, Sep. 2015.
[20] D. Maji, M. A. Suster, E. Stavrou, U. A. Gurkan, and P. Mohseni, “Monitoring time course of human whole blood coagulation using a microfluidic dielectric sensor with a 3D capacitive structure,” in Proc. 37th Annu. Int. IEEE Eng. Med. Biol. Conf., pp. 5904–5907, Aug. 2015.
[21] http://home.ku.edu.tr/~ozlemyalcin/aggregation1.html
[22] J.-J. Luo, S.-J. Huang, C.-L. Yang, “Portable Power-Saving Blood Coagulation Detection Devices for Remote Health Care Systems,” in Biomedical Engineering Society Annual Symposium, Dec. 2009.
[23] Keysight Technologies, "Basics of Measuring the Dielectric Properties of Materials," Published in USA, May 16, 2014.
[24] O. K. Baskurt, M. Uyuklu, and H. J. Meiselman, “Time course of electrical impedance during red blood cell aggregation in a glass tube: comparison with light transmittance,” Biomedical Engineering, IEEE Transactions on, vol. 57, pp. 969-978, 2010.
[25] 邱彥超, 高準確度之可攜式血液凝固檢測儀之設計及其臨床驗證, 國立成功大學,台南, 台灣, 2012.
[26] A. Albishi and O. M. Ramahi, “Detection of surface and subsurface cracks in metallic and non-metallic materials using a complementary split-ring resonator,” Sensors, vol. 14, no. 10, pp. 19 354–19 370, 2014.
[27] A. Ebrahimi, W. Withayachumnankul, S. Al-Sarawi, and D. Abbott, “High-sensitivity metamaterial-inspired sensor for microfluidic dielectric characterization,” IEEE Sensors J., vol. 14, no. 5, pp. 1345–1351, May 2014.
[28] L. Benkhaoua, M. T. Benhabiles, S. Mouissat, and M. L. Riabi, “Miniaturized Quasi-Lumped Resonator for Dielectric Characterization of Liquid Mixtures,” IEEE Sensors Journal, vol. 16, pp. 1603-1610, 2016.
[29] R. Melik, E. Unal, N. Perkgoz, B. Santoni, D. Kamstock, C. Puttlitz, and H. Demir, “Nested metamaterials for wireless strain sensing,” IEEE J. Sel. Topics Quantum Electron., vol. 16, no. 2, 2010.
[30] Y. Feldman, I. Ermolina, and Y. Hayashi, “Time domain dielectric spectroscopy study of biological systems,” IEEE Trans. Dielect. Elect. Insulation, vol. 10, pp. 728–753, Oct. 2003.
[31] 謝谷孟, 應用環型指叉電極進行糖化血色素阻抗量測晶片之研究, 國立成功大學,台南, 台灣, 2013.
[32] A. Velez, F. Aznar, J. Bonache, M. C. Velázquez-Ahumada, J. Martel, and F. Martín, “Open complementary split ring resonators (OCSRRs) and their application to wideband CPW band pass filters,” IEEE Microw. Wireless Compon. Lett., vol. 19, no. 4, pp. 197–199, Apr. 2009.
[33] S. Karthikeyan and R. Kshetrimayum, "Compact, harmonic suppressed power divider using open complementary split‐ring resonator," Microwave and Optical Technology Letters, vol. 53, pp. 2897-2899, 2011.
[34] Keysight Technologies, "E4990A Impedance Analyzer Data Sheet," Published in USA, August 5, 2016.
[35] T. Nakatsuka, "Polylactic acid-coated cable," Fujikura Tech. Rev, vol. 3946, 2011.
[36] 周承緯, 新型可攜式光學凝血偵測儀於全血凝血酶原時間檢測之評估, 國立成功大學,台南, 台灣, 2012.
[37] 許家華, 應用自動平衡橋於電阻抗法凝血時間檢測之電路設計與實現, 國立成功大學, 台南, 台灣, 2014.