| 研究生: |
郭瑞明 Guo, Rei-Ming |
|---|---|
| 論文名稱: |
以模印高分子薄膜修飾電極用於對臨床尿液檢體之肌酸酐交流阻抗式感測 Preparation of the imprinted polymeric film modified electrode for AC impedance sensing of creatinine in clinical urine specimens |
| 指導教授: |
黃世宏
Huang, Shyh-Hong |
| 學位類別: |
碩士 Master |
| 系所名稱: |
工學院 - 化學工程學系 Department of Chemical Engineering |
| 論文出版年: | 2019 |
| 畢業學年度: | 107 |
| 語文別: | 中文 |
| 論文頁數: | 75 |
| 中文關鍵詞: | 肌酸酐 、模印高分子 、生物感測器 、電化學 、交流阻抗頻譜 、尿液臨床檢體 |
| 外文關鍵詞: | creatinine, molecularly imprinted polymers, electrochemical sensor, electrochemical impedance spectroscopy, urine specimens |
| 相關次數: | 點閱:112 下載:0 |
| 分享至: |
| 查詢本校圖書館目錄 查詢臺灣博碩士論文知識加值系統 勘誤回報 |
近年來我國邁入高齡化社會,國人對衛生保健與食品安全更加重視,而生物感測器的發展日漸蓬勃,應用亦日漸廣闊,其快速檢測、高靈敏度、可攜式且操作簡便等特點吸引許多研究人員將其應用於臨床檢驗與居家照護等領域。肌酸酐 (Creatinine) 為肌酸 (creatine) 與磷酸肌酸 (creatine phosphate) 之代謝廢物,是人類血液與尿液中用來評估腎臟功能的重要指標。
本研究以肌酸酐作為模版分子,NVP (N-vinylpyrrolidone) 以及NVI (1-vinylimidazole) 作為功能性單體進行共聚合形成高分子薄膜,利用萃洗過程將肌酸酐從高分子中洗脫,產生與肌酸酐具有特異性吸附效果之模印高分子 (molecularly imprinted polymer, MIP) 薄膜電極。
本研究以 FT-IR 之官能基特徵峰存在與否驗證肌酸酐成功地從高分子膜洗脫出;並以 SEM 分析各階段修飾電極的表面型態;以循環伏安法 (cyclic voltammetry) 探討各階段修飾電極電化學活性面積的變化,並以 EIS 數據模擬出等效電路;透過感測環境與電極修飾條件的最適化,並利用交流阻抗頻譜對肌酸酐作定量檢測,其濃度對應阻抗變化率在10~120 mg/dL呈高度線性,即使在標準尿中於10~300 mg/dL的範圍亦呈高度線性。本研究開發之肌酸酐感測器對人類尿液檢體進行感測與比對,由檢體測量值可確認此生物感測器於臨床檢驗的可行性。
SUMMARY
Creatinine, the final product of the metabolism of creatine in mammals that removed from the body by kidneys. It is a reliable indicator for kidney’s function in human’s blood and urine. Biosensors are already well established in modern analytical chemistry. They have become important tools for clinical diagnostics, environmental analysis, production monitoring, drug detection or screening. Synthetic biomimetic receptors like molecularly imprinted polymers (MIPs) have shown to be a potential alternative to biomolecules as recognition element for biosensing.
In this work, sensors based on poly(N-vinylpyrrolidone-co-1-vinylimidazole) for the selective determination of creatinine are fabricated by heated polymerization, We not only discuss the conclusion of creatinine in MIP by FTIR characterization and morphology, but also optimize creatinine calibration and detection by electrochemical impedance spectroscopy (EIS). Finally, the fabricated creatinine sensor was applied to the detection of urine specimens for feasibility test on clinical application.
Keywords: creatinine, molecularly imprinted polymers, electrochemical sensor, electrochemical impedance spectroscopy, urine specimens
Introduction
Creatinine, is the dehydrogenated form of creatine in ATP metabolism. The normal range of serum creatinine is 0.5~1.5 mg/dL; while in urine, the normal range is 39~259 mg/dL for males and 28~217 mg/dL for females, respectively. The value varies with human’s age, lifestyle and muscle mass. The concentration of creatinine is less affected by the variation of dietary, therefore creatinine in blood and urine is a fairly reliable indicator for clinical evaluation of renal function.
Clinical analysis of creatinine is routinely carried out using colorimetric Jaffé reaction and enzymatical reaction. However, Jaffé reaction suffers from the poor selectivity toward numerous metabolites containing carbonyl group found, such as proteins, glucose, bilirubin, and ascorbic acid. Although the introduction of enzymes may improve specificity, enzymes always suffer from instability, high cost and complex procedures for the immobilization.
In recent years, molecularly imprinted polymers (MIPs) have been used to develop the new generation of biological/chemical sensors due to their superiorities such as high adsorption capacity, high specific recognition of selected analyte, low cost and excellent reusability.
In this work, N-vinylpyrrolidone and 1-vinylimidazole as functional monomers were copolymerized in the presence of creatinine by heat polymerization. Based on the formation of hydrogen bonds of these functional monomers and creatinine, the fabricated electrode can be introduced to improve the specific binding affinity of creatinine. By the linearity between impedance change and creatinine concentration, we can apply in urine specimen detection.
MATERIALS AND METHODS
The gold working electrode was prepared by a sputter coater (sputter coater 108 auto, Cressington). Calculation of the amount of creatinine removed is measured by HPLC (SPD-10A; SIL-9A; C-R6A). Incorporation of creatinine in MIP matrix before and after extraction was compared by FTIR. Morphology of the modified electrodes were observed by SEM images. All electrochemical experiments were performed by using (Electrochemical analyzer) AUT85707 and Zive SP1 controlled and acquired by electrochemical signals of the sample solution. Clinical urine specimens were collected from NCKU hospital for detection test by the MIP modified electrodes.
RESULTS AND DISCUSSION
The proportion of creatinine removed from the MIP film was measured by an HPLC and calculated as 80.2%, which showed good imprinting as well as removal efficiency. FTIR analysis for the MIP films before and after template removal, the peaks at 3,250 cm1 which is amide group (NH stretching) disappeared. Meanwhile, the CN group at 1,108 cm1 decreased, which showed creatinine was removed from the polymer.
From the EIS (electrochemical impedance spectroscopy) simulation of modified electrode, the mechanism of creatinine detection can be confirmed. Because of formation of cavities inside the polymeric film, fascilitating the electron transfer between background electrolyte and working electrode, then with the injection of creatinine solution, creatinine can be rebinded to recognition sites in the cavities, blocked the electron transfer and resulted in the increase of impedance of whole electrode.
The as-prepared MIP electrode was examined by electrochemical detection. The gold electrode was modified by allyl mercaptan and imprinted poly(N-vinylpyrrolidone-co-1-vinylimidazole) was synthesized onto the electrode with the optimal ratio (NVP:NVI= 4:1), The sensing background was optimized at pH 7.4 to get the best calibration curve. Linear calibration curve with a slope of 0.024 ·mg1dL was thus obtained with R2= 0.9971. By comparison between MIP and NIP (Non-molecularly imprinted polymer), the imprinting factor was 12, the modified electrode showed the specific binding towards creatinine molecules. Real urine sample test was accomplished by the modified electrode and HPLC, respectively. Comparison to these methods, the result is good enough for clinical detection and it preliminary feasibility in point of care (POC) detection.
The sensing performance of the MIP electrode in this work such as reusability and storage stability was also investigated, discussing the reusability and storage stability of this MIP electrode. After reused for 50 times, the MIP electrode still remains remarkable stability compared to the same electrode prepared in the very beginning.
Conclusion
The extraction process to remove creatinine from the polymer matrix is confirmed from HPLC measurement and FTIR analysis. Both the preparation of MIP electrode and detection of creatinine were optimized. The as-prepared MIP electrode can detect urine creatinine within the range of 10~120 mg/dL versus the electrochemical impedance change with a sensitivity of 0.024 ·mg1dL and a LOD of 5.68 mg/dL. Additionally, the MIP electrode thus prepared can be applied to the determination of creatinine concentration in urine specimens.
1. S Wang, T Chinnasamy, M Lifson, F Inci, U Demirci. Flexible substrate-based devices for point-of-care diagnostics. Trends in Biotechnology, 34, 909921, 2016.
2. J Wang. Glucose biosensors: 40 years of advances and challenges. Electroanalysis, 13, 983‒988, 2001.
3. FH Cincotto, EL Fava, FC Moraes, O Fatibello-Filho, RC Faria. A new disposable microfluidic electrochemical paper-based device for the simultaneous determination of clinical biomarkers. Talanta, 195, 6268, 2019.
4. MU Ahmed, MM Hossain, M Safavieh, YL Wong, IA Rahman, M Zourob, E Tamiya. Toward the development of smart and low cost point-of-care biosensors based on screen printed electrodes. Critical Reviews in Biotechnology, 36, 495‒505, 2016.
5. Z Zhou, D Du, J Wang, JN Smith, C Timchalk, Y Li, Y Lin. Quantum dot-based immunochromatographic fluorescent biosensor for biomonitoring trichloropyridinol, a biomarker of exposure to chlorpyrifos. Analytical Chemistry, 82, 5125‒5133, 2010.
6. 教育部顧問室「生物技術科技教育改進計畫」分子檢驗。2003
7. S Lee, H Jo, J Her, H Lee, C Ban. Ultrasensitive electrochemical detection of engrailed-2 based on homeodomain-specific DNA probe recognition for the diagnosis of prostate cancer. Biosensors and Bioelectronics, 66, 32‒38, 2015.
8. M Subat, AS Borovik, B König. Synthetic creatinine receptor: imprinting of a Lewis acidic zinc(II)cyclen binding site to shape its molecular recognition selectivity. Journal of American Chemical Society, 126, 3185‒3190, 2004.
9. T Guinovart, DH Alonso, L Adriaenssens, P Blondeau, FX Rius, P Ballester, FJ Andrade. Characterization of a new ionophore-based ion-selective electrode for the potentiometric determination of creatinine in urine. Biosensors and Bioelectronics, 87, 587‒592, 2017.
10. S Song, L Wang, L Wang, J Li, J Chao, C Fan. Aptamer-based biosensors. Trac-Trends in Analytical Chemistry, 27, 108‒117, 2008.
11. N Aydemir, J Malmstrom, J Travas-Sejdic. Conducting polymer based electrochemical biosensors. Physical Chemistry Chemical Physics, 18, 8264‒8277, 2016.
12. F Wei, S Cheng, Y Korin, EF Reed, D Gjertson, C Ho, HA Gritsch, J Veale. Serum creatinine detection by a conducting-polymer-based electrochemical sensor to identify allograft dysfunction. Analytical Chemistry, 84, 7933‒7937, 2012.
13. A Diouf, S Motia, NE Hassani, NE Bari, B Bouchikhi. Development and characterization of an electrochemical biosensor for creatinine detection in human urine based on functional molecularly imprinted polymer. Journal of Electroanalytical Chemistry, 788, 4453, 2017.
14. GM Holzinger, AL Gof, S Cosnier. Supramolecular immobilization of bio-entities for bioelectrochemical applications. New Journal of Chemistry, 38, 5173-5180, 2014.
15. J Ballesta-Claver, J Ametis-Cabello, J Morales-Sanfrutos, A Megía-Fernández, MC Valencia-Mirón, F Santoyo-González, LF Capitán-Vallvey. Electrochemiluminescent disposable cholesterol biosensor based on avidin–biotin assembling with the electroformed luminescent conducting polymer poly(luminol-biotinylated pyrrole). Analytica Chemica Acta, 91‒98, 2012.
16. A Zamora-Galvez, E Morales-Narvaez, CC Mayorga-Martinez, A Merkoci. Nanomaterials connected to antibodies and molecularly imprinted polymers as bio/receptors for bio/sensor applications. Applied Materials Today, 9, 387‒401, 2017.
17. M Dabrowski, M Cieplak, PS Sharma, P Borowicz, K Noworyta, W Lisowski, F D'Souza, A Kuhn. Hierarchical templating in deposition of semi-covalently imprinted inverse opal polythiophene film for femtomolar determination of human serum albumin. Biosensors and Bioelectronics, 94, 155‒161, 2017.
18. R Schirhagl. Bioapplications for Molecularly Imprinted Polymers. Analytical Chemistry, 86, 250‒261, 2014.
19. H Tsai, C Lin, Y Juang, I Wang, Y Lin, R Wang, H Lin. Multiple type biosensors fabricated using the CMOS BioMEMS platform. Sensors and Actuators B, 144, 407412, 2010.
20. M Syu, T Hsu, Z Lin. Synthesis of recognition matrix from 4-methylamino-N-allylnaphthal-imide with fluorescent effect for the imprinting of creatinine. Analytical Chemistry, 82, 8821‒8829, 2010.
21. J Lee. Bioluminescence, the Nature of the Light. University of Georgia Libraries, 2017.
22. Y Yanase, T Hiragun, K Ishii, T Kawaguchi, T Yanase, M Kawai, K Sakamoto, M Hide. Surface plasmon resonance for cell-based clinical diagnosis. Sensors, 14, 4948‒4959, 2014.
23. JC Wang, WH Lin, E Cao, XF Xu, WJ Liang, XF Zhang. Surface plasmon resonance sensors on raman and fluorescence spectroscopy. Sensors, 17, 2719, 2017.
24. S Unser, I Bruzas, J He, L Sagle. Localized surface plasmon resonance biosensing: current challenges and approaches. Sensors, 15, 15684‒15716, 2015.
25. https://en.wikipedia.org/wiki/Raman_spectroscopy.
26. R Ahmed, AK Yetisen, SH Yun, H Butt. Color-selective holographic retroreflector array for sensing applications. Light-Science & Applications, 6, e16214, 2017.
27. M Elsherif, MU Hassan, AK Yetisen, H Butt. Glucose sensing with phenylboronic acid functionalized hydrogel-based optical diffusers. ACS Nano, 12, 2283‒2291, 2018.
28. D Zhang, Q Liu. Biosensors and bioelectronics on smartphone for portable biochemical detection. Biosensors and Bioelectronics, 75, 2016.
29. D Capoferri, R Álvarez-Diduk, M Del Carlo, D Compagnone, A Merkoçi. Electrochromic molecular imprinting sensor for visual and smartphone-based detections. Analytical Chemistry, 90, 5850‒5856, 2018.
30. S Zhang, R Geryak, J Geldmeier, S Kim, VV Tsukruk. Synthesis, assembly, and applications of hybrid nanostructures for biosensing. Chemical Reviews, 117, 12942‒13038, 2017.
31. R Gui, H Jin, H Guo, Z Wang. Recent advances and future prospects in molecularly imprinted polymers-based electrochemical biosensors. Biosensors and Bioelectronics, 100, 56‒70, 2018.
32. Y Fuchs, O Soppera, K Haupt. Photopolymerization and photostructuring of molecularly imprinted polymers for sensor applications—A review. Analytica Chemica Acta, 717, 7‒20, 2012.
33. G Wulff, J Haarer. The preparation of defined chiral cavities for the racemic resolution of free sugars. Markromol.Chem, 192, 1329‒1338, 1991.
34. H Rao, Z Lu, H Ge, X Liu, B Chen, P Zou, X Wang, H He, X Zeng, Y Wang. Electrochemical creatinine sensor based on a glassy carbon electrode modified with a molecularly imprinted polymer and a Ni@polyaniline nanocomposite. Microchim Acta, 184, 261‒269, 2017.
35. B Babamiri, A Salimi, R Hallaj, M Hasanzadeh. Nickel nanoclusters as a novel emitter for molecularly imprinted electrochemiluminescence based sensor toward nanomolar detection of creatinine. Biosensors and Bioelectronics, 107, 272279, 2018.
36. L N Gómez-Arribas, J L Urraca, E Benito-Peña, M C Moreno-Bondi. Tag-Specific Affinity Purification of Recombinant Proteins by Using Molecularly Imprinted Polymers. Analytical Chemistry, 91, 4100‒4106, 2019.
37. 柯定賢。合成螢光單體以製備具螢光之模版高分子膜用於螢光式感測肌酸酐。國立成功大學化工系碩士學位論文,2009
38. X Zhang, F Zheng, L Ye, P Xiong, L Yan, W Yang, B Jiang. A one-pot sol-gel process to prepare a superhydrophobic and environment-resistant thin film from ORMOSIL nanoparticles. RSC Advances, 4, 9838‒9841, 2014.
39. J Coresh, BC Astor, T Greene, G Eknoyan, AS Levey. Prevalence of chronic kidney disease and decreased kidney function in the adult US population: Third National Health and Nutrition Examination Survey. American Journal of Kidney Diseases, 41, 1‒12, 2003.
40. 吳明儒。評估腎臟功能的方法。腎臟語病析,19,45-49,2007
41. SR Dunn, GM Gabuzda, KR Superdock, RS Kolechi, RW Schaedler, ML Simenhoff. Induction of creatininase activity in chronic renal failure: Timing of creatinine degradation and effect of antibiotics. American Journal of Kidney Diseases, 29, 7277, 1997.
42. M Peake, M Whiting. Measurement of Serum Creatinine – Current Status and Future Goals. The clinical biochemist reviews, 4, 173‒184, 2006.
43. V Nguyen, C Wolff, J Seris, J Schwing. Immobilized Enzyme Electrode for Creatinine Determination in Serum. Analytical Chemistry, 63, 611614, 1991.
44. F Wei, S Cheng, Y Korin, E Reed, D Gjertson, C Ho, H Gritsch, J Veale. Serum Creatinine Detection by a Conducting Polymer Based Electrochemical Sensor to Identify Allograft Dysfunction. Analytical Chemistry, 84, 7933‒7937, 2012.
45. EP Randviir, CE Banks. Analytical methods for quantifying creatinine within biological media. Sensors and Actuators B: Chemical, 183, 239252, 2013.
46. EP Randviir, DK Kampouris, CE Banks. An improved electrochemical creatinine detection method via a Jaffé-based procedure. Analyst, 138, 6565‒6572, 2013.
47. AJ Killard, MR Smyth. Creatinine biosensors: principles and designs. Trends in Biotechnology, 18, 433‒437, 2000.
48. T Wen, W Zhu, C Xue, J Wu, Q Han, X Wang, X Zhou, H Jiang. Novel electrochemical sensing platform based on magnetic field-induced self-assembly of Fe3O4@Polyaniline nanoparticles for clinical detection of creatinine. Biosensors and Bioelectronics, 56, 180‒185, 2014.
49. Q Ang, F Chan, P Tan, S Low. Improving imprinted shape cavities of molecularly imprinted sol–gel host matrix with minimal relaxation for sensing of creatinine. Journal of Sol-Gel Science and Technology, 86, 226238,2018.
50. AJ Bard, LR Faulkner. Electrochemical methods: fundamentals and applications. 2001.
51. J Huang, Z Li, H Ge, JB Zhang. Analytical solution to the impedance of electrode/electrolyte interface in lithium-ion batteries. Journal of the electrochemical Society, 162, A7037A7048, 2015.
52. N Elgrishi, KJ Rountree, BD McCarthy, ES Rountree, TT Eisenhart, JL Dempsey. A practical beginner’s guide to cyclic voltammetry. Journal of Chemical Education, 95, 197‒206, 2018.
53. F Scholz, AM Bond, RG Compton, DA Fiedler, G Inzelt, H Kahlert. Electroanalytical methods: guide to experiments and applications. 2nd Ed., Springer, Inc., Berlin 2010.
54. D Pihíková, P Kasák, J Tkac. Glycoprofiling of cancer biomarkers: Label-free electrochemical lectin-based biosensors. Open Chemistry, 13, 636‒655, 2015.
55. E Barsoukov, JR Macdonald. Impedance spectroscopy theory, experiment, and applications. 3rd Ed., John Wailey& Sons, Inc., USA 2018.
56. JB Jorcin, ME Orazem, N Pébère, B Tribollet. CPE analysis by local electrochemical impedance spectroscopy. Electrochemica Acta, 51, 1473‒1479, 2006.
57. 許庭榕。以分子模版技術用於螢光式與電化學阻抗式分析感測肌酸酐。國立成功大學化工系碩士學位論文,2010
58. 黃宥霖。製備溶膠凝膠模版高分子薄膜電極用於肌酸酐濃度之交流阻抗式感測。國立成功大學化工系碩士學位論文,2015
59. KK Reddy, KV Gobi. Artificial molecular recognition material based biosensor for creatinine by electrochemical impedance analysis. Sensors and Actuators B: Chemical, 183, 356363, 2013.
60. B Khadro, C Sanglar, A Bonhomme, A Errachid, N Jaffrezic-Renault. Molecularly imprinted polymers (MIP) based electrochemical sensor for detection of urea and creatinine. Procedia Engineering, 5, 371374, 2010.
61. C Huang, T Tsai, J Thomas, M Lee, B Liu, H Lin. Urinalysis with molecularly imprinted poly(ethylene-co-vinyl alcohol) potentiostat sensors. Biosensors and Bioelectronics, 24, 26112617, 2009.
62. F Martinez-Pina, L Gargallo, D Radic. N-vinylimidazole-N-vinyl-2-pyrrolidone copolymers. Part I. reactivity ratios, solubility and viscosimetric study. Polymer International, 47, 340344, 1998.
63. S Ida, H Kitanaka, T Ishikawa, S Kanaoka, Y Hirokawa. Swelling properties of thermoresponsive/hydrophilic co-networks with functional crosslinked domain structures. Polymer Chemistry, 9, 17011709, 2018.
64. https://en.wikipedia.org/wiki/1-Vinylimidazole#cite_note-12
65. J Du, B Zhu, WR Leow, S Chen, TC Sum, X Peng, X Chen. Colorimetric detection of creatinine based on plasmonic nanoparticles via synergistic coordination chemistry. Small, 33, 4104‒4110, 2015.
66. CM Stumpe, H Grubmueller. Aqueous urea solutions: structure, energetics, and urea aggregation. Journal of Physical Chemistry B, 111, 62206228, 2007.
校內:2029-06-25公開